This study aims at assessing the mechanical behavior of a composite metal/polymer bar to be implanted in the retrosternal position, in order to correct chest wall deformities, such as Pectus Excavatum. A 300-mm-long, 12.7-mm-wide, and 3.5-mm-thick Nuss bar was considered, made of different metals and biodegradable polymers, fixed at its extremities, and with a constant force of 250 N applied on its center. Two different geometries for the metal elements to be embedded in the polymeric matrix were tested: in the former, thin metal sheets and in the latter, cylindrical metal reinforcing rods were considered. Finite element method simulation results are reported, in terms of maximum stress and strain of the bar. Furthermore, the maximum stress values obtained by varying metal sheet thickness or rod diameter (and therefore the volumetric percentage of metal within the matrix) for different material combinations are also shown; optimal configuration for the Pectus Excavatum treatment was finally identified for a composite Nuss bar.

Metal/polymer composite Nuss bar for minimally invasive bar removal after Pectus Excavatum treatment: FEM simulations

RICOTTI, Leonardo;CIUTI, GASTONE;MENCIASSI, Arianna
2014-01-01

Abstract

This study aims at assessing the mechanical behavior of a composite metal/polymer bar to be implanted in the retrosternal position, in order to correct chest wall deformities, such as Pectus Excavatum. A 300-mm-long, 12.7-mm-wide, and 3.5-mm-thick Nuss bar was considered, made of different metals and biodegradable polymers, fixed at its extremities, and with a constant force of 250 N applied on its center. Two different geometries for the metal elements to be embedded in the polymeric matrix were tested: in the former, thin metal sheets and in the latter, cylindrical metal reinforcing rods were considered. Finite element method simulation results are reported, in terms of maximum stress and strain of the bar. Furthermore, the maximum stress values obtained by varying metal sheet thickness or rod diameter (and therefore the volumetric percentage of metal within the matrix) for different material combinations are also shown; optimal configuration for the Pectus Excavatum treatment was finally identified for a composite Nuss bar.
2014
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11382/492776
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 2
social impact