The study aim was to evaluate and compare the analytical performance of the new chemiluminescent immunoassay for cardiac troponin I (cTnI), called Access hs-TnI using DxI platform, with those of Access AccuTnI+3 method, and high-sensitivity (hs) cTnI method for ARCHITECT platform. The limits of blank (LoB), detection (LoD) and quantitation (LoQ) at 10% and 20% CV were evaluated according to international standardized protocols. For the evaluation of analytical performance and comparison of cTnI results, both heparinized plasma samples, collected from healthy subjects and patients with cardiac diseases, and quality control samples distributed in external quality assessment programs were used. LoB, LoD and LoQ at 20% and 10% CV values of the Access hs-cTnI method were 0.6, 1.3, 2.1 and 5.3 ng/L, respectively. Access hs-cTnI method showed analytical performance significantly better than that of Access AccuTnI+3 method and similar results to those of hs ARCHITECT cTnI method. Moreover, the cTnI concentrations measured with Access hs-cTnI method showed close linear regressions with both Access AccuTnI+3 and ARCHITECT hs-cTnI methods, although there were systematic differences between these methods. There was no difference between cTnI values measured by Access hs-cTnI in heparinized plasma and serum samples, whereas there was a significant difference between cTnI values, respectively measured in EDTA and heparin plasma samples. Access hs-cTnI has analytical sensitivity parameters significantly improved compared to Access AccuTnI+3 method and is similar to those of the high-sensitivity method using ARCHITECT platform.

Evaluation of analytical performance of a new high-sensitivity immunoassay for cardiac troponin i

Masotti, Silvia;MUSETTI, VERONICA;Passino, Claudio;Clerico, Aldo
2018-01-01

Abstract

The study aim was to evaluate and compare the analytical performance of the new chemiluminescent immunoassay for cardiac troponin I (cTnI), called Access hs-TnI using DxI platform, with those of Access AccuTnI+3 method, and high-sensitivity (hs) cTnI method for ARCHITECT platform. The limits of blank (LoB), detection (LoD) and quantitation (LoQ) at 10% and 20% CV were evaluated according to international standardized protocols. For the evaluation of analytical performance and comparison of cTnI results, both heparinized plasma samples, collected from healthy subjects and patients with cardiac diseases, and quality control samples distributed in external quality assessment programs were used. LoB, LoD and LoQ at 20% and 10% CV values of the Access hs-cTnI method were 0.6, 1.3, 2.1 and 5.3 ng/L, respectively. Access hs-cTnI method showed analytical performance significantly better than that of Access AccuTnI+3 method and similar results to those of hs ARCHITECT cTnI method. Moreover, the cTnI concentrations measured with Access hs-cTnI method showed close linear regressions with both Access AccuTnI+3 and ARCHITECT hs-cTnI methods, although there were systematic differences between these methods. There was no difference between cTnI values measured by Access hs-cTnI in heparinized plasma and serum samples, whereas there was a significant difference between cTnI values, respectively measured in EDTA and heparin plasma samples. Access hs-cTnI has analytical sensitivity parameters significantly improved compared to Access AccuTnI+3 method and is similar to those of the high-sensitivity method using ARCHITECT platform.
2018
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11382/520628
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 52
social impact